Industry Applications
Applies to HDI, FPC, chip packaging substrate, PCB auxiliary materials, etc.
Product Introduction
1. Flexible, multi-channel and group test capability
Establish multi groups of test modes to test different samples in 1-4 environmental test chambers according to the test demands.
2. Wide-range bias voltage and test voltage
The test voltage includes 300V, 500V and 1000V, and can be set randomly within 1.0V ~ 1000V, and the positive and negative bias voltage can be switched during the test.
3. Real-time current detection capability
During the test, the 10 times/s micro-current monitoring module can draw the working status of ionic migration process between channels for material analysis.
4. Threshold judgment of multiple insulation values
Support multiple judging criteria of insulation resistance threshold, for example, the test will be disconnected after single or multiple triggering of insulation resistance threshold, or lower than threshold running time.
5. Test system software developed for domestic clients
Fully conform to the special test requirements of different clients. Fitted with data analysis function to make the test software simpler and better.
6. International and industrial standards of insulation resistance test
IPC-TM-650 2.5.10/2.6.11/2.6.3/2.6.14 IEC-61189-5 IEC1086 IS0-9455-17, etc.
7. Remote functions of insulation resistance system
Support remote connection and control of Windows system desktop through the platform of Win, Android, iOS and Mac, to realize remote control and viewing of computer and operation of test software.
System Parameters
|
Model |
300V |
500V |
1000V |
|
Number of channels |
32 /64 |
||
|
Number of test groups |
1/2 groups |
1/2 groups |
1 group |
|
Working hours |
1~9,999 hours (adjustable) |
||
|
Bias voltage |
-100~+300VDC |
-100~+500VDC |
-100~+1,000VDC |
|
Polarity inversion |
Available |
||
|
Common-pole test |
Common positive pole |
||
|
CAF test (optional) |
Available |
||
|
Test voltage |
1.0~100VDC (0.01V stepping) |
||
|
>100VDC (0.1V stepping) |
|||
|
Voltage rise cascade |
100V (>100V) |
||
|
Precision of test voltage |
±0.1V (1.0~100VDC) |
||
|
±0.5V (>100VDC) |
|||
|
Test scope of resistance |
1x105~3x1013Ω |
1x105~5x1013Ω |
1x105~1x1014Ω |
|
Resistance test precision |
1x106~1x1010Ω≤±5% (Test voltage 5V) |
1x106~1x1010Ω≤±5% (Test voltage 5V) |
1x106~1x1010Ω≤±5% (Test voltage 5V) |
|
1x1010~1x1011Ω≤±10% (Test voltage 5V) |
1x1010~1x1011Ω≤±10% (Test voltage 5V) |
1x1010~1x1011Ω≤±10% (Test voltage 5V) |
|
|
1xl011~lx1012Ω≤±15% (Test voltage 50V) |
1x1011~lx1012Ω≤±15% (Test voltage 50V) |
1x1011~lx1012Ω≤±15% (Test voltage 50V) |
|
|
1x1012~1x1013Ω≤±20% (Test voltage 300V) |
1x1012~1x1013Ω≤±20% (Test voltage 500V) |
1x1012~1x1013Ω≤±20% (Test voltage 1,000V) |
|
|
Resistance test speed |
3.5s/channel (typical) |
||
|
64 channels <5min |
|||
|
Scope of real-time current monitoring |
0.1 ~300μA |
0.1~500μA |
0.1~1,000μA |
|
Speed of real-time current detection |
10 times/channel |
||
|
Test interval |
5min~600min (adjustable) |
||
|
Voltage rise cascade |
1s~600s (adjustable) |
||
|
Protection resistance |
1MΩ |
||
|
Test cable |
Teflon (>1014Ω, -55~200℃); standard configuration: 3.0m |
||
|
Time of continuous power supply by UPS |
300x400x500 (mm) (HxWxD) |
||
|
Overall dimensions |
About 30kg |
About 30kg |
About 35kg |
|
Input voltage |
AC 220V±10%, 50Hz |
||
Search
BACK
Consultation Hotline